New TopShim run:

version = topshim 2.0.0
date = 2023-10-27
time = 16:02:53.155
command = topshim  convcomp tuneaxyz

1D SHIMMING

Parameters:

maximum order = 5
probe = Z116098_0395
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 7.24 ppm
optimisation parameters = ls
linewidth 1H = 0.10 Hz
envelope shape / strictness = 2.00 / 200

Results:

tune lock level increase = 1.3

initial B0 stdDev = 2.95 Hz
sample size = 2.89 cm, position = 0.02 cm
final B0 stdDev = 0.32 Hz > improvement = 9.3
envelope width = 0.71 Hz

Performing tune after shimming
tune lock level increase = 1.0

shim changes: 
Z	+54
Z2	-39
Z3	+98
Z4	-249
Z5	+72
X	-52
Y	+123

duration = 3 min 37 sec
completed successfully
finished Fri Oct 27 16:06:28 2023


========================================================================